월 | 화 | 수 | 목 | 금 | 토 | 일 |
---|---|---|---|---|---|---|
09:00 - 17:00 | 09:00 - 17:00 | 09:00 - 17:00 | 09:00 - 17:00 | 09:00 - 17:00 | X | X |
의뢰는 파우더 샘플만 가능 합니다 |
X-선 회절원리를 이용한 물질의 결정구조 분석기 무기/고분자/제약 나노물질과 박막의 결정구조 분석을 통하여 물질확인 및 물질변화를 정밀하게 측정
1. X-ray generator
1) Max. output : 600W or more
2) Max. voltage : 40kV or more
3) Max. current : 15mA or more
4) X-ray tube : Cu tube
5) Safety features
2. Radiation Enclosure
3. Goniometer
1) Geometry : Vertical θ/2θ
2) Drive axis : θ/2θ coupled (mechanical linked axis)
3) Driving method : single axis regular interval control
4) Scan range : -3° to +145°(2θ) or wider
5) Goniometer radius:150mm or longer
6) Accuracy : ±0.02° or better
4. Slit system
1) Divergence slit : 0.625°, 1.25° or better
2) Variable DS
- continuous variable mode : X-ray irradiation width is kept constant
at approx. 20mm in use below approx. 67°(2θ)
3) Anti-scattering slit : 1.25° or better
4) Receiving slit : 0.3mm or better
5) Soller slit : 2.5° or 4°
5. 1D high speed detector
1) Mainbody : 1D semiconductor strip type detector
2) Effective element detection area : 256mm² (12.8 x 20mm) or bigger
3) Energy resolution : 25% or less (@CuKa)
4) Counting efficiency : More than 98% (@CuKa)
5) Knife edge unit : reducing scattering background at low angle area.
(1) Diffraction angle variable type or equivalent
6. Automated sample changer
1) Max. No. of samples : Six(6) samples
2) Sample rotation is provided
표시되지 않은 시료전처리, 재료비 및 분석결과의 특별한 해석을 의뢰할 경우에 별도로 요금을 산정함
구분 | 항목 | 단위 | 외부 | 교내 | 가족회사 | 성의교정 | 비고 |
---|---|---|---|---|---|---|---|
분석의뢰 |
분석료 | 샘플 | 30,000 | 18,000 | 24,000 | 24,000 | |
전처리 | 샘플 | 10,000 | 6,000 | 8,000 | 8,000 | ||
직접사용 |
분석료 | 30분당 | 15,000 |